Philips Azurion R3.1 Software Recall
Philips Azurion system; Software Version Number: R3.1;
Updated December 12, 2025
Philips Azurion systems running Software Version R3.1 have two software issues that can cause loss of X-ray imaging, loss of motorized movement, or incorrect images. One issue may trap the system in continuous restart mode; the other affects systems with a Poly-G3 frontal stand and causes positioning errors.
Is the PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalled? Yes. This product was recalled by FDA on December 12, 2025. Details below.
What is recalled
Philips Azurion system with Software Version Number R3.1, manufactured by Philips Medical Systems Nederland B.V.
Am I affected
You're affected if you operate or own a Philips Azurion system running Software Version R3.1. Check your system's software version in the system settings. If your system has a Poly-G3 frontal stand, you may also experience longitudinal positioning errors.
What to do right now
Contact Philips Medical Systems immediately for a software update or further instructions. Do not use the system if it enters continuous restart mode or if imaging functionality is lost until you receive guidance from Philips.
The hazard
Philips has identified two (2) software issues affecting device systems that may result in loss of imaging (X-ray) functionality and/or loss of motorized movement, and/or incorrect image content and/or loss of data. Issue 1 - System remains in continuous restart mode after the start-up. Issue 2 - Longitudinal Position Error Applicable only to systems with Poly-G3 frontal stand.
Brands affected
- Source
- FDA
- Category
- Medical Devices
- Recall ID
- fda-Z-1113-2026
- Classification
- Class II
- Recalled
- December 12, 2025
Get free recall alerts
New Medical Devices recalls, the moment they post.
Free forever. Unsubscribe anytime. We never sell your data.
More Medical Devices recalls
Boston Scientific Rapid Refill Injection System Recall
Boston Scientific's Rapid Refill Continuous Injection System contains a recalled syringe component that may loosen or disconnect from the manifold. This can …
Boston Scientific Corporation
FARASTAR Ablation Generator Recall
Boston Scientific is replacing certain FARASTAR Pulsed Field Ablation generators that cannot work with the FARAPOINT catheter. If your generator has an …
Boston Scientific Corporation
Faradive Steerable Sheath Clear Recall
The Faradive Steerable Sheath Clear used in cardiac catheterization procedures may allow air to enter the bloodstream if the dilator is inserted incorrectly …
Boston Scientific Corporation