Medline Dialysis Dressing Change Kit Recall
Convenience kits containing Tego Connectors Medline DIALYSIS DRSNG CHNG SKU DYNDC3167
Updated January 8, 2026
Medline dialysis dressing change convenience kits with Tego Connectors may have defective silicone seals that can tear or dome, blocking the fluid path and potentially exposing patients to biological contaminants or air infusion during treatment.
Is the Medline Industries, LP recalled? Yes. This product was recalled by FDA on January 8, 2026. Details below.
What is recalled
Convenience kits containing Tego Connectors, Medline Dialysis Dressing Change SKU DYNDC3167.
Am I affected
You're affected if you use Medline dialysis dressing change convenience kits with Tego Connectors (SKU DYNDC3167) that show visible silicone seal damage, doming, or tearing on the connector.
What to do right now
Stop using affected kits immediately. Contact Medline Industries, LP to report the issue and obtain a replacement kit. Do not attempt to use a kit with a damaged or compromised Tego Connector seal.
The hazard
Affected lots may exhibit issues with the silicone seal on Tego Connectors included in kits. Issues include silicone seal doming and silicone seal tearing. These issues may result in an occluded fluid path identified by the inability of the user to inject or withdraw blood product with a syringe. Damaged seals may also result in delay in therapy or fluid leakage, interruption in therapy, exposure to biological contaminants or air infused into the body.
Brands affected
- Source
- FDA
- Category
- Medical Devices
- Recall ID
- fda-Z-1468-2026
- Classification
- Class II
- Recalled
- January 8, 2026
Get free recall alerts
New Medical Devices recalls, the moment they post.
Free forever. Unsubscribe anytime. We never sell your data.
More Medical Devices recalls
Boston Scientific Rapid Refill Injection System Recall
Boston Scientific's Rapid Refill Continuous Injection System contains a recalled syringe component that may loosen or disconnect from the manifold. This can …
Boston Scientific Corporation
FARASTAR Ablation Generator Recall
Boston Scientific is replacing certain FARASTAR Pulsed Field Ablation generators that cannot work with the FARAPOINT catheter. If your generator has an …
Boston Scientific Corporation
Faradive Steerable Sheath Clear Recall
The Faradive Steerable Sheath Clear used in cardiac catheterization procedures may allow air to enter the bloodstream if the dilator is inserted incorrectly …
Boston Scientific Corporation