Da Vinci S, Si Cautery Hook Instrument Recall
Brand Name: da Vinci S, Si Permanent Cautery Hook Instrument Product Name: da Vinci S, Si Permanent Cautery Hook Instrument Model/Catalog Nu
Updated December 9, 2025
Intuitive Surgical's da Vinci S and Si Permanent Cautery Hook Instruments (model 420183) may have frayed or broken pitch cables that transmit motion to the surgical tool tip. A cable failure during surgery could compromise the instrument's function and patient safety.
Is the Intuitive Surgical, Inc. recalled? Yes. This product was recalled by FDA on December 9, 2025. Details below.
What is recalled
Da Vinci S and Si Permanent Cautery Hook Instruments (Model 420183). These are reusable surgical instruments with mechanical tungsten cables used in minimally invasive endoscopic surgery.
Am I affected
You are affected if you own or use a da Vinci S or Si system with the Permanent Cautery Hook Instrument, model 420183. Check your instrument's model number and serial documentation.
What to do right now
Contact Intuitive Surgical, Inc. immediately to report any instruments showing frayed or broken pitch cables. Do not use affected instruments until the issue is addressed. Consult your system documentation or contact your service representative for guidance on inspection and replacement options.
The hazard
Due to increased in complaints regarding frayed or broken pitch cables on reusable surgical instruments
Brands affected
- Source
- FDA
- Category
- Medical Devices
- Recall ID
- fda-Z-1701-2026
- Classification
- Class II
- Recalled
- December 9, 2025
Get free recall alerts
New Medical Devices recalls, the moment they post.
Free forever. Unsubscribe anytime. We never sell your data.
More Medical Devices recalls
Boston Scientific Rapid Refill Injection System Recall
Boston Scientific's Rapid Refill Continuous Injection System contains a recalled syringe component that may loosen or disconnect from the manifold. This can …
Boston Scientific Corporation
FARASTAR Ablation Generator Recall
Boston Scientific is replacing certain FARASTAR Pulsed Field Ablation generators that cannot work with the FARAPOINT catheter. If your generator has an …
Boston Scientific Corporation
Faradive Steerable Sheath Clear Recall
The Faradive Steerable Sheath Clear used in cardiac catheterization procedures may allow air to enter the bloodstream if the dilator is inserted incorrectly …
Boston Scientific Corporation