Gentuity HF-OCT Imaging System Recall
Gentuity HF-OCT Imaging System Model G10-01; Software version 23.2.13;
Updated March 23, 2026
The Gentuity HF-OCT Imaging System Model G10-01 with certain software versions can produce repeated frames during imaging pullbacks, which may lead to inaccurate length measurements in optical coherence tomography scans. Gentuity is working with users to ensure they understand how to recognize and avoid this issue.
Is the Gentuity, LLC recalled? Yes. This product was recalled by FDA on March 23, 2026. Details below.
What is recalled
Gentuity HF-OCT Imaging System Model G10-01 with software version 23.2.13, when used with Vis-Rx and Vis-Rx Prime catheters and Gentuity HF-OCT Console software versions 21.11 to 23.3.13.
Am I affected
You're affected if you operate a Gentuity HF-OCT Imaging System Model G10-01 with software version 23.2.13, and use Vis-Rx or Vis-Rx Prime catheters with HF-OCT Console software version 21.11 through 23.3.13.
What to do right now
Contact Gentuity, LLC directly for updated software and guidance on how to recognize and mitigate repeated frames during pullback procedures to ensure accurate longitudinal measurements.
The hazard
The potential for repeated frames during HF-OCT pullback when using the Gentuity HF-OCT Console (software version 21.11 to 23.3.13) with the Vis-Rx and Vis-Rx Prime catheters. This correction is intended to ensure that users are fully informed of how to recognize, avoid, and mitigate this condition to maintain accurate longitudinal (length) measurements during OCT imaging.
Brands affected
- Source
- FDA
- Category
- Medical Devices
- Recall ID
- fda-Z-1823-2026
- Classification
- Class II
- Recalled
- March 23, 2026
Get free recall alerts
New Medical Devices recalls, the moment they post.
Free forever. Unsubscribe anytime. We never sell your data.
More Medical Devices recalls
Boston Scientific Rapid Refill Injection System Recall
Boston Scientific's Rapid Refill Continuous Injection System contains a recalled syringe component that may loosen or disconnect from the manifold. This can …
Boston Scientific Corporation
FARASTAR Ablation Generator Recall
Boston Scientific is replacing certain FARASTAR Pulsed Field Ablation generators that cannot work with the FARAPOINT catheter. If your generator has an …
Boston Scientific Corporation
Faradive Steerable Sheath Clear Recall
The Faradive Steerable Sheath Clear used in cardiac catheterization procedures may allow air to enter the bloodstream if the dilator is inserted incorrectly …
Boston Scientific Corporation