GE Allia IGS 3 Pulse X-ray System Recall
Allia IGS 3 Pulse angiographic X-ray system
Updated April 20, 2026
An electrical cable defect in certain GE Allia IGS 3 Pulse angiographic X-ray systems equipped with Gaia generators poses a risk of electrical shock to service personnel during installation, maintenance, or troubleshooting when the system is powered on.
Is the GE Medical Systems, LLC recalled? Yes. This product was recalled by FDA on April 20, 2026. Details below.
What is recalled
GE HealthCare Allia IGS 3 Pulse angiographic X-ray systems equipped with Gaia generators.
Am I affected
You are affected if you are a service technician or personnel who installs, maintains, or troubleshoots GE Allia IGS 3 Pulse systems with Gaia generators, particularly when working inside the C-FRT cabinet with power on.
What to do right now
Contact GE HealthCare immediately if you service or maintain one of these systems. Do not open or work inside the C-FRT cabinet while the system is powered on until you have received guidance from the manufacturer.
The hazard
GE HealthCare has become aware of a potential issue affecting an electrical cable on certain IGS systems equipped with Gaia generators. This issue could lead to a risk of electrical shock to service personnel performing installation, maintenance, or troubleshooting activities inside the C-FRT cabinet while power is ON.
Brands affected
- Source
- FDA
- Category
- Medical Devices
- Recall ID
- fda-Z-2314-2026
- Classification
- Class II
- Recalled
- April 20, 2026
Get free recall alerts
New Medical Devices recalls, the moment they post.
Free forever. Unsubscribe anytime. We never sell your data.
More Medical Devices recalls
Boston Scientific Rapid Refill Injection System Recall
Boston Scientific's Rapid Refill Continuous Injection System contains a recalled syringe component that may loosen or disconnect from the manifold. This can …
Boston Scientific Corporation
FARASTAR Ablation Generator Recall
Boston Scientific is replacing certain FARASTAR Pulsed Field Ablation generators that cannot work with the FARAPOINT catheter. If your generator has an …
Boston Scientific Corporation
Faradive Steerable Sheath Clear Recall
The Faradive Steerable Sheath Clear used in cardiac catheterization procedures may allow air to enter the bloodstream if the dilator is inserted incorrectly …
Boston Scientific Corporation